Skip to main content
Peptides TT home

Glossary term

HPLC (High-Performance Liquid Chromatography)

HPLC (high-performance liquid chromatography) is an analytical technique that separates the components of a sample on a packed column so that the purity of a peptide can be measured.

Also known as: High-pressure liquid chromatography, RP-HPLC

High-performance liquid chromatography (HPLC) is the main technique used to measure the purity of a synthetic peptide. A small sample load is introduced into a stream of solvent that is pumped at high pressure through a column packed with fine particles. Different molecules interact with the packing to different degrees, so they leave the column at different times and can be detected one after another.

How peptide HPLC works

Most peptide analysis uses reversed-phase HPLC. The column is packed with silica particles carrying C18 or C8 hydrocarbon chains, and the mobile phase is a mixture of water and acetonitrile, usually containing about 0.1% trifluoroacetic acid. Over the run, the proportion of acetonitrile is increased in a gradient. More hydrophobic species stay on the column longer and elute later.

A UV detector monitors the column outflow, typically at 214 or 220 nm, where the peptide bond absorbs strongly. The output is a chromatogram: a trace of detector response against time, with each separated component appearing as a peak.

Reading the chromatogram

  • Retention time is the time at which a peak elutes. Under fixed conditions it is characteristic of a given compound.
  • Peak area is proportional to the amount of UV-absorbing material in that peak.
  • Area percent purity is the area of the main peak divided by the total area of all integrated peaks, expressed as a percentage.

Small peaks either side of the main peak usually represent closely related impurities, such as deletion sequences, truncated chains or oxidised forms.

Why it matters in the lab

HPLC gives a reproducible, quantitative picture of how much of a batch is the intended peptide. It does have limits. An impurity that co-elutes with the main peak can be hidden, and HPLC on its own does not prove that the main peak is the correct sequence. That is why it is paired with mass spectrometry, which confirms identity by measuring molecular mass. Aggregation-prone sequences such as Amyloid Beta 1-42 can also need adjusted conditions to give a clean profile.

The HPLC trace and purity figure for each batch appear on its certificate of analysis. Our article on peptide purity testing by HPLC and MS walks through example chromatograms in more detail.